Eqiupment: Instrumentum GC (Shimadzu GC-2010)
Columna: DB-17 Agilent 30mX0.53mmX1.0μm
Clibano temperatus initialis: 80℃
Initial time 2.0min
Rate 15℃/min
Clibano temperatus finalis: 250℃
Final tempus 20min
Portitorem gas Nitrogen
Modus Constans fluxus
Fluere 5.0mL/min
Split ratio 10:1
Injector temperatus: 250℃
Detector temperatus: 300℃
Iniectio volumen 1.0μL
Cavendi sunt ante analysis:
1. Conditio columnae ad 240℃ pro minimis 30minutis.
2. clysterem lava et liner injectoris munda proprie ad contaminantes analyseos priorum removendos.
3. Lava, sicca et dilue in clystere imple phialas lava.
Praeparatio diluens;
Praeparet 2% w/v solutionem sodium hydroxidum in aqua.
Praeparatio sacra:
Expende circa 100mg of (R) hydroxyprrolidinum hydrochloridi-3-vexillum in phialam, 1mL diluentium et dissolve adde.
Praeparatio Test:
Expende circa 100mg specimen experimenti in phialam, adde 1mL diluentium et dissolve.Para in du- plici.
De modo procedendi:
Inject blank (diluente), vexillum praeparatio et probatio super conditionibus GC utens.Neglegere iugis debitum ad blank.Retentionis tempus apicem propter (R) -3-hydroxyprrolidine circiter 5.0min est.
Nota:
Referre effectus in medium
Post tempus: Nov-13-2021